8. Setting of parameters
Page 96
105204
– WTS-300-E - V4.1 - 20.01.2022
Section 8.5
: Level “Extended cleaning parameters 1” (r parameters)
Settings cleaning breaks
Range
Def.
r70 Activation of break in cleaning programme (see section 11.2)
0
0: deactivated
1: activated, finish of break only with button or digital input
2: activated, finish of break after time [r72] or button or digital input
3: activated, finish of break at a certain time [r73] or button or digital input
r71 Position break
After which cleaning stage is the break integrated
Setting range: after cleaning stage 0….8
0 .. 8
1
r72 Break time
0...999 min.
120
r73 Finish of the break at time
00:00..23:59 00:00
Settings of cleaning pump (see section 11.9)
Range
Def.
r80 Cleaning pump venting
0 .. 4
0
0: deactivated
1: On-/off-impulses with a fixed quantity for all cleaning stages
2: as 1, however, premature termination if pressure via pressure switch
3: as 1, however, unlimited impulses when rinsing (cleaning stages 0+9)
4: as 2, however, unlimited impulses when rinsing (cleaning stages 0+9)
r81 Number of on / off impulses cleaning pump
0 .. 10
r82 ON impulse length cleaning pump
0 .. 10 sec.
5
r83 OFF impulse length cleaning pump
0 .. 10 sec.
5
Settings for rinsing cycle-optimisation
Range
Def.
r86 Water pressure-controlled rinse-out step.
If the water pressure in the rinse-out step (n15,
n25...n85) drops out for x seconds, the rinse se-
quence is continued with the drain step (n16,
n26...n86).
0: deactivated
1..6: activated, seconds without pressure.
0..6 sec.
0
r87 Avoidance of rinsing shadows on the agitator
blade
0: deactivated
1: in the circulation phase (n14, n24, n34...n84)
the agitator stops every 25 seconds for 5 sec.
0..1
0
Selection 1, 2, 3:
If the digital input is to be
used to terminate the
pause, the function 15
must additionally be as-
signed to one of the digi-
tal inputs A21-A28 in the
A- parameters.