dPMR Signal Rx Meter Remote Commands
4 - 16
4.8
SIGNAL POWER
4.8.1
Signal Power - Averages
:METERs:POWer:CH1:AVERaging
:METERs:POWer:CH1:AVERaging?
4.8.2
Signal Power - Average Measurement Reset
:METERs:POWer:CH1:CLEAR:AVG
4.8.3
Signal Power - Lower Limit Enable
:LIMits:POWer:CH1:LOWer:ENABle
:LIMits:POWer:CH1:LOWer:ENABle?
Description:
Set command defines the number of readings taken to calculate Average Signal
Power measurement.
Query command returns parameter setting.
Range:
1 to 250
Default Value:
1
Set/Query Format:
NR1
Example:
:METERs:POWer:CH1:AVERaging 100
Sets the number of readings taken to calculate Average Signal Power
measurement to 100.
Query Response:
:METERs:POWer:CH1:AVERaging?
100
Description:
Command clears and resets Average Signal Power measurement.
Parameter/Query:
none
Description:
Set command Enables/Disables Lower Limit for Signal Power measurement.
Query command returns parameter setting.
Parameter:
OFF | ON | 0 | 1
Default Value:
OFF
Set/Query Format:
Boolean
Example:
:LIMits:POWer:CH1:LOWer:ENABle ON
Enables Lower Limit for Signal Power measurement.
Query Response:
:LIMits:POWer:CH1:LOWer:ENABle?
1
Содержание 3900 Series
Страница 1: ...3900 Series Digital Radio Test Set dPMR Remote Programming Manual ...
Страница 4: ...Preface ii THIS PAGE INTENTIONALLY LEFT BLANK ...
Страница 6: ...Contents iv THIS PAGE INTENTIONALLY LEFT BLANK ...
Страница 44: ...Generator Transmit Channel Remote Commands 2 34 THIS PAGE INTENTIONALLY LEFT BLANK ...
Страница 69: ...Analyzer Receiver Remote Commands 3 25 THIS PAGE INTENTIONALLY LEFT BLANK ...
Страница 147: ...Audio Demod Signal Rx Meter Remote Commands 5 54 THIS PAGE INTENTIONALLY LEFT BLANK ...