Generator/Transmit Channel Remote Commands
2 - 3
2.2.5
AF Generators - Waveform
:AF:GENerator:SOURceN:SHAPe
:AF:GENerator:SOURceN:SHAPe?
Description:
Set command defines the Waveform for the specified AF Generator.
Query command returns parameter setting.
Parameter:
SINE | SQUare | TRIangle | RAMP | DCS | DCSINV | DTMF
Query Data:
SNR | SINE | SQUare | TRIangle | RAMP | DCS | DCSINV | DTMF | TONESEQ |
TONEREM
Default Value:
SINE
Set/Query Format:
CPD | CRD
Example:
:AF:GENerator:SOURce2:SHAPe SQUare
Sets AF Generator 2 Waveform shape to Square.
Query Response:
:AF:GENerator:SOURce2:SHAPe?
SQU
SourceN = 1, 2 or 3 (AF Generator 1, 2 or 3)
DTMF waveform is only valid on AF Generator 1. AF Generator 2 is unavailable
when DTMF is selected on AF Generator 1.
DCS and DCSINV are not supported on AF Generator 3.
AF Generator 1 is unavailable as a modulation source when Normal MOD SNR
Noise Measurements are defined (:CONFigure:MOD:ANALyzer:SNR:MODE 1)
and AF:GENerator:SOURce1:SHAPe? returns SNR.