Audio/Demod Signal Rx Meter Remote Commands
5 - 43
5.16.5
Mod Hum & Noise - Lower Limit Value
:LIMits:MOD:HN:LOWer:VALue
:LIMits:MOD:HN:LOWer:VALue?
5.16.6
Mod Hum & Noise - Measurement Query
:FETCh:MOD:ANALyzer:HN?
5.16.7
Mod Hum & Noise - Peak Measurement Reset
:MOD:ANALyzer:HN:CLEAR:PEAK
5.16.8
Mod Hum & Noise - Reference Lock
:CONFigure:MOD:ANALyzer:HN:REFerence
Description:
Set command defines Lower Limit Value for Mod Hum & Noise measurement.
Query command returns parameter setting.
Range:
-100.0 to +100.0 dBr
Units:
dBr
Default Value:
0.0 dBr
Set/Query Format:
NRf | NR1
Example:
:LIMits:MOD:HN:LOWer:VALue -50dBr
Sets Lower Limit Value for Mod Hum & Noise measurement to -50.0 dBr.
Query Response:
:LIMits:MOD:HN:LOWer:VALue?
-50
Description:
Command returns Mod and Noise measurement data.
Query Data:
<statusbyte>,<failbyte>,<avgcount>,<avg(dB)>,<wc(dB)>
Refer to Chapter 1 for UUT Measurement Meter Return Data.
Query Response:
:FETCh:MOD:ANALyzer:HN?
4,4,13,62.35,75.68
Mod Analyzer Noise measurement type must be defined as Signal to Noise
Ratio (:MOD:ANALyzer:NTYPe SN) and SNR measurement defined as Hum &
Noise (:CONFigure:MOD:ANALyzer:SNR:MODE 0) to obtain valid Hum & Noise
measurement.
Description:
Command clears and resets Peak Mod Hum & Noise measurement.
Parameter/Query:
none
Description:
Command locks Mod Hum and Noise reference to current meter reading.
Parameter/Query:
none
SNR measurement must be defined as Hum & Noise to obtain valid Hum &
Noise measurement (:CONFigure:MOD:ANALyzer:SNR:MODE HN).
N O T E
N O T E
N O T E
Содержание 3900 Series
Страница 1: ...3900 Series Digital Radio Test Set dPMR Remote Programming Manual ...
Страница 4: ...Preface ii THIS PAGE INTENTIONALLY LEFT BLANK ...
Страница 6: ...Contents iv THIS PAGE INTENTIONALLY LEFT BLANK ...
Страница 44: ...Generator Transmit Channel Remote Commands 2 34 THIS PAGE INTENTIONALLY LEFT BLANK ...
Страница 69: ...Analyzer Receiver Remote Commands 3 25 THIS PAGE INTENTIONALLY LEFT BLANK ...
Страница 147: ...Audio Demod Signal Rx Meter Remote Commands 5 54 THIS PAGE INTENTIONALLY LEFT BLANK ...