ProStar325 Service Manual
64
Publication date 8 May 2003
•
The replacement procedure is the opposite of this. Use care
to align the flowcell on locating pins.
3.21.1
Tubing
Each flowcell comes with a set of nuts & ferrules that are fitted to
peek tubing and connected to the flowcell as shown.
Part
number
Description
Used on all
types except
4x0.15
16 101265 00
Fitting 1/16 tube nut/ferrule flat
bottom
(Pack of 10)
Used on
flowcell 4x0.15
only
16 101269 00
16 101264 00
Fitting 1/8 tube nut flat bottom
(each)
Fitting 1/8 tube ferrule (Pack of 10)
The flowcells are constructed from titanium and are provided
with quartz optics windows. Tubing connections are PEEK type
except for the Super Prep flowcell which uses TEFZEL TUBE
0.125 x 0.062.
Each flowcell is easily interchangeable by the user.
thumbscrew
StockCheck.com
Содержание ProStar 325 UV-Vis
Страница 9: ...ProStar325 Service Manual Publication date 8 May 2003 9 S t o c k C h e c k c o m ...
Страница 19: ...ProStar325 Service Manual Publication date 8 May 2003 19 S t o c k C h e c k c o m ...
Страница 34: ...ProStar325 Service Manual 34 Publication date 8 May 2003 S t o c k C h e c k c o m ...
Страница 71: ...ProStar325 Service Manual Publication date 8 May 2003 71 S t o c k C h e c k c o m ...
Страница 72: ...ProStar325 Service Manual 72 Publication date 8 May 2003 S t o c k C h e c k c o m ...
Страница 99: ...ProStar325 Service Manual Publication date 8 May 2003 99 Deuteriu m lamp S t o c k C h e c k c o m ...
Страница 118: ...ProStar325 Service Manual 118 Publication date 8 May 2003 S t o c k C h e c k c o m ...
Страница 149: ...ProStar325 Service Manual Publication date 8 May 2003 149 S t o c k C h e c k c o m ...