ProStar325 Service Manual
132
Publication date 8 May 2003
7.10.1
Input/Output
To activate this page you must be in Diagnostic State.
Go to 'Diagnostic'> 'Main' and select ‘i/o collect mode’. Then tick
'Enter Diagnostic State'.
Note: you must first select i/o collect mode before entering into
the diagnostic state or else this page does not become active.
Analog Supply Voltages -
Used to view power supply voltage
levels coming from the two rear power supplies PS1 & PS2. These
measurements are performed by the ADC circuits on the main
Processor PWB.
Analog In
- Verify the Analog input and output circuits (DACA
& DACB) by setting the sliders on the two Analog Out controls to
any value and then have this voltage measured and displayed in
the corresponding Analog In (DACA & DACB) fields. This can be
used to verify chart recorder and external integrator connections.
StockCheck.com
Содержание ProStar 325 UV-Vis
Страница 9: ...ProStar325 Service Manual Publication date 8 May 2003 9 S t o c k C h e c k c o m ...
Страница 19: ...ProStar325 Service Manual Publication date 8 May 2003 19 S t o c k C h e c k c o m ...
Страница 34: ...ProStar325 Service Manual 34 Publication date 8 May 2003 S t o c k C h e c k c o m ...
Страница 71: ...ProStar325 Service Manual Publication date 8 May 2003 71 S t o c k C h e c k c o m ...
Страница 72: ...ProStar325 Service Manual 72 Publication date 8 May 2003 S t o c k C h e c k c o m ...
Страница 99: ...ProStar325 Service Manual Publication date 8 May 2003 99 Deuteriu m lamp S t o c k C h e c k c o m ...
Страница 118: ...ProStar325 Service Manual 118 Publication date 8 May 2003 S t o c k C h e c k c o m ...
Страница 149: ...ProStar325 Service Manual Publication date 8 May 2003 149 S t o c k C h e c k c o m ...