ProStar325 Service Manual
106
Publication date 8 May 2003
required for removing the Processor PWB).
•
On the Processor PWB, disconnect P13, P17, P22 & P15.
•
Carefully lift the optics module out of the detector and
place safely on a flat bench.
•
Remove the two screws on the large silver clamp that
secures the black foam padding around the optics module.
•
Take off all the padding pieces to expose the
monochromator. This also enables access for replacing the
Dual Photodiodes Detector PWB.