Note:
Always check the information printed on the rear of the
detector for the most up-to-date details regarding the fuse type.
To check a fuse -
1/. Disconnect the detector from the mains power supply.
2/. On the mains input connector (rear panel), lift the fuse cap
and pull out the fuse holder (lower red section).
3/. Both fuses are held in this red fuse holder.
4/. Check that the fuses are the correct type and are not
damaged. If necessary, replace them.
5/. Place the fuses into the red fuse holder, push the holder back
into the mains input connector and then lower the cap.
6/. Reconnect the instrument to the mains power supply.
WARNING:
FIRE HAZARD
To avoid fire hazard replace fuses ONLY with fuses of
the same rating and type. Always switch off the AC
power and unplug the power cord before replacing a
fuse.
To expose the fuse compartment, lift the cap using a small flat
blade screwdriver.
Mains
input
connector
Fuse
compartme
nt
StockCheck.com
Содержание ProStar 325 UV-Vis
Страница 9: ...ProStar325 Service Manual Publication date 8 May 2003 9 S t o c k C h e c k c o m ...
Страница 19: ...ProStar325 Service Manual Publication date 8 May 2003 19 S t o c k C h e c k c o m ...
Страница 34: ...ProStar325 Service Manual 34 Publication date 8 May 2003 S t o c k C h e c k c o m ...
Страница 71: ...ProStar325 Service Manual Publication date 8 May 2003 71 S t o c k C h e c k c o m ...
Страница 72: ...ProStar325 Service Manual 72 Publication date 8 May 2003 S t o c k C h e c k c o m ...
Страница 99: ...ProStar325 Service Manual Publication date 8 May 2003 99 Deuteriu m lamp S t o c k C h e c k c o m ...
Страница 118: ...ProStar325 Service Manual 118 Publication date 8 May 2003 S t o c k C h e c k c o m ...
Страница 149: ...ProStar325 Service Manual Publication date 8 May 2003 149 S t o c k C h e c k c o m ...