ProStar325 Service Manual
146
Publication date 8 May 2003
Calibrate with D2 lamp on
Calibrate with both lamps on
These will be required under the following circumstances;
Calibrate with
Vis lamp on
Calibrate with
D2 lamp on
Calibrate with
both lamps on
Fitting Vis lamp option
Y
Y
Y
Replacing Vis lamp
Y
Y
Y
Replacing UV lamp
if Vis fitted
Y
if Vis fitted
Replacing Processor
PWB
if Vis fitted
Y
if Vis fitted
Replacing Detector
PWB
if Vis fitted
Y
if Vis fitted
Replacing mono
if Vis fitted
Y
if Vis fitted
Refractive Index test
fails
if Vis fitted
Y
if Vis fitted
Detector Linearity test
fails
if Vis fitted
Y
if Vis fitted
Establish communications as described in the procedure above.
Turn the D2 Lamp ON by clicking the D2 Lamp button, and wait
until it is ready.
Both the ’Ready’ and ’D2 Lamp is ON’ indicators should be green.
StockCheck.com
Содержание ProStar 325 UV-Vis
Страница 9: ...ProStar325 Service Manual Publication date 8 May 2003 9 S t o c k C h e c k c o m ...
Страница 19: ...ProStar325 Service Manual Publication date 8 May 2003 19 S t o c k C h e c k c o m ...
Страница 34: ...ProStar325 Service Manual 34 Publication date 8 May 2003 S t o c k C h e c k c o m ...
Страница 71: ...ProStar325 Service Manual Publication date 8 May 2003 71 S t o c k C h e c k c o m ...
Страница 72: ...ProStar325 Service Manual 72 Publication date 8 May 2003 S t o c k C h e c k c o m ...
Страница 99: ...ProStar325 Service Manual Publication date 8 May 2003 99 Deuteriu m lamp S t o c k C h e c k c o m ...
Страница 118: ...ProStar325 Service Manual 118 Publication date 8 May 2003 S t o c k C h e c k c o m ...
Страница 149: ...ProStar325 Service Manual Publication date 8 May 2003 149 S t o c k C h e c k c o m ...