ProStar325 Service Manual
Publication date 8 May 2003
75
mode Q8 is regulating the current and the voltage feedback is
from the D2 lamp’s cathode via D15.
Warning
If the polycarbonate safety cover is removed, the D2
supply
voltage is exposed. Contact with this voltage
may cause
electric shock or death. Minimize this
risk by keeping the cover
in place.
The following is the sequence of events when the D2 lamp
striking;
Start
Set the current
to the filament
Wait 10seconds
Is the
filament
on?
Strike lamp and wait
0.5seconds
Has the
lamp
struck?
Disable filament and
strike controls
No
N
Yes
Yes