ProStar325 Service Manual
Publication date 8 May 2003
89
•
Locate and remove the two screws on the rear panel
•
Remove the optics module as described in the Optics section
of this manual.
•
Remove the earth lug screw located on the Processor PWB.
•
Remove the single screw holding down the power supply
module. This is located on the outer edge of the detector
adjacent to the power supply module.
•
Disengage the safety cover and slide the power supply
module toward the front of the detector. At this stage the
StockCheck.com