56
4.4
Output format
Output format of measured data are as follows
■
Data format
No. Data
Unit
Data example
Details
1
Range for OPEN filter
-
1
Measurement range for OPEN filter
2
Range for X2 filter
-
2
Measurement range for X2 filter
3
Range for Y filter
-
3
Measurement range for Y filter
4
Range for Z filter
-
8
Measurement range for Z filter
5
Count
-
54321
A/D count value
6
Equivalent voltage
V
2.65
Equivalent voltage for A/D count value
7
Correction factor number
-
0
Current correction factor number
8
Luminance
cd/
㎡
3.4567E+001
Luminance
9
Tristimulus values X
-
4.5678E+001
Tristimulus values X
10
Tristimulus values Y
cd/
㎡
3.4567E+001
Tristimulus values Y
11
Tristimulus values Z
-
2.3456E+001
Tristimulus values Z
12
Chromaticity x
-
0.2345
Chromaticity x
13
Chromaticity y
-
0.3456
Chromaticity y
14
Chromaticity
u’
-
0.1234
Chromaticity
u’
15
Chromaticity
v’
-
0.5678
Chromaticity
v’
16
Color Temperature
K
2856
Color Temperature
17
Deviation
-
0.0456
Deviation
18
Data termination
-
“END”
Termination in output data
※
1 The instrument send
“
****
” for out of valid range.
※
2 Output data for Items that is not in active use Items are
”
****
”
※
3 The data of Count and Equivalent voltage are sent only when set single filter.