33
3.2.4.2
Entry correction factor
The procedure for entry correction factor is as follows;
1
Select following screen in the function mode.
2
Press [CHANGE] switch and following screen appear.
3
Press [ROTATION] switch to move cursor [
*
] to [INPUT].
4
Press [ENTER] switch and following screen appears.
Press [ROTATION] switch to select correction factor data from recorded data.
Example we enter correction factor number to be
“01”.
(
R=NEXT
:
[ROTATION]
、
S=PAGE
:
[SHIFT]
)
5
Press [CHANGE] switch and following menu screen appears.
-
DIRECT
Select this item to enter correction factor KX
、
KY
、
KZ directly.
-
REF. & MEASURE
Select this item to enter reference luminance/chromaticity data, and measure the object in order
to calculate correction factor in the RD-80SA.
-
BACK
FACTOR
DISP&SET OFF
INPUT *BACK
CLEAR
* FACTOR NUMBER *
SET FACTOR OFF
FACTOR
DISP&SET OFF
*INPUT BACK
CLEAR
K01
NO DATA
R=NEXT S=PAGE
FACTOR INPUT
DIRECT
REF. & MEASURE
*BACK