46
4.3.6
SRL Command
This command instructs the instrument to send serial number of the instrument
Valid value range
:
00000000
– 99999999
PC
Command RD-80SA
“SRL”+CR+LF
“OK”+CR+LF
“12CR+LF
“END”+CR+LF
4.3.7
ST Command
The instrument instructs the instrument to start measurement immediately after receive this
command according to specified each setting. When finish measurement, the instrument sends
measured data according to
「
4.4
Output format of Measurement data
」
-
Sending the command from PC to the instrument
PC or PLC send character string
“ST” (0x53 0x54)
+
CR(0x0d)
+
LF(0x0a) via LAN or RS-232C
-
Return from the instrument to the PC
The instrument returns
“OK+CR+LF” immediately and starts measurement after the instrument
receive
”ST” command.
The instrument sends measured data to PC according to
「
4.4 Output format of Measurement
data
」
after finish measurement. If error occur in measurement, the instruments sends
”NG”
+
CR
+
LF to PC. You can check the details of err by sending
”ERR”
+
CR
+
LF to the instrument.
PC
Command RD-80SA
“ST”+CR+LF
“OK”+CR+LF
Start measurement
・
・
・
・
・
Complete measurement
DATA1+CR+LF
DATA2+CR+LF
・
・
CR+LF
“END”+CR+LF