43
FX_#
Sets measurement range for X2 filter in [RANGE EACH]
Valid value range
:
1 - 8
FY_#
Sets measurement range for Y filter in [RANGE EACH]
Valid value range
:
1
– 8
FZ_#
Sets measurement range for Z filter in [RANGE EACH]
Valid value range
:
1 - 8
MF_#
Selects filter
0
:
OPEN
1
:
X2 filter
2
:
Y filter
3
:
Z filter
4
:
X2,Y,Z filter
FS_#
Specifies the correction factor number to be enabled.
Valid value range
:
0 - 10
※
0:correction fact is disabled
FG
Reads current correction factor number.
WF_#_####_####_####
Enters correction factor data.
Valid value range
:
Correction factor number
:
1 -10
Correction factor
:
0.01
– 100.0
RF_#
Reads correction factor data in specifying correction
factor number
Valid value range
:
1 - 10
CF_#
Initializes specifying correction factor.
When specify
“0”, All correction factor data are initialized.
Valid value range
:
0 - 10
BR_#
Sets communication speed in RS-232C
0
:
9600 1
:
19200
2
:
38400
DM_#
Sets delimiter in RS-232C
0
:
CR+LF 1
:
CR
LIS_###.###.###.###
Sets IP address in LAN
Setting takes active after restart the instrument.
Valid value range
:
192.168.100.1 - 192.168.255.255
LIG
Reads LAN IP address.
LMS_###.###.###.###
Sets sub net mask in LAN
Setting takes active after restart the instrument.
Valid value range
:
255.255.0.0 - 255.255.255.255
LMG
Reads sub net mask in LAN
LPS_#####
Sets LAN port number
Setting takes active after restart the instrument.
Valid value range
:
49152 - 65535
LPG
Reads LAN port number
PMT
Sets sensitivity mode of photo detector
0
:
INITIAL
1
:
USER
ERR
Reads latest error number.
”
_
” means Space. “#” means numerical values.
Delimiter CR
:
0x0d
、
LF
:
0x0a