48
4.3.9
SPC Command
This command stops the continuous measurement triggered by the STC command. The instrument
will send the status of continuous measurement.
Return value
0
:
Normal stop
1
:
Continuous measurement
2
:
Abnormal stop
PC
Command RD-80SA
“SPC”+CR+LF
“OK”+CR+LF
Stop measurement
・
・
・
・
・
“0”+CR+LF
“END”+CR+LF
※
The instrument do not send
“END” command until completing first measurement.
4.3.10
SMS Command
Reads the latest status of continuous measurement. This command is available only when the
instrument is in continuous measurement mode.
Return value
0
:
Normal stop
1
:
Continuous measurement
2
:
Abnormal stop
PC
Command RD-80SA
“SMS”+CR+LF
“OK”+CR+LF
“2”+CR+LF
“END”+CR+LF
※
When instrument receive this command, the instrument does not send
“END” command until
completing all measurement.
4.3.11
SMD Command
Reads latest measured data in continuous measurement
PC
Command RD-80SA
“SMD”+CR+LF
“OK”+CR+LF
Measured data
・
・
・
・
・
“END”+CR+LF