36
Measured data of object
8
Press [ENTER] switch to return Correction factor menu screen.
3.2.4.3
Check and set of correction factor
Entered correction factor shows. When correction factor is not entered, the RD-80SA shows [NO
DATA]. [SET] appears on the screen in applied correction factor.
No correction factor entry
Correction factor data
Applied correction factor
Press [CHANGE] switch to determine or cancel.
Details of items
K01
:
Correction factor Number. The RD-80SA can record ten correction factor data.
K01 - K10
NO DATA
:
When no correction factor is entered, this message appear.
KX / KY / KZ
:
Correction factor in Tristimulus values.
C=SET
:
[CHANGE] switch
=
SET
Press [CHANGE] switch to apply the value on the screen to correction factor.
[SET] appears on the screen in applied correction factor.
R=NEXT
:
[ROTATION] switch
=
NEXT
Another correction factor appears.
S=PAGE
:
[SHIFT] switch
=
PAGE
Each time pressing [SHIFT] switch, screen change as follows;
Reference data
→ Measured data → Correction factor data → Reference data
SMP01 x= 0.4464
y= 0.4075
L= 9.980E+01
R=NEXT S=PAGE
K01 KX= 1.000E+00
KY= 1.000E+00
KZ= 1.000E+00
C=SET R=NEXT S=PAGE
K01
NO DATA
R=NEXT S=PAGE
K01 KX= 1.000E+00
SET KY= 1.000E+00
KZ= 1.000E+00
C=SET R=NEXT S=PAGE