60
Measurement range
(for standard illuminant A)
About 0.1 cd/m
2
-
10000 cd/m
2
RANGE
Measurement range
(
cd/
㎡)
1
0.1
-
5
2
0.5
-
15
3
1.5
-
40
4
3.5
-
120
5
15
-
600
6
35
-
1600
7
220
-
2900
8
750
-
10000
※
Output voltage of 1V at range 5 is adjusted to 200cd/m
2
for standard illuminant A.
※
3.4V is upper limit of output voltage in each range.
※
Above value is design reference value. Measurable range may vary from the performance of
density filter.
Linearity (for Standard illuminant A, Auto range)
Luminance
:
±3%
2 cd/
㎡
or below
±2%
2 cd/
㎡
or above
Accuracy is guaranteed for measurable range only
Chromaticity (for standard illuminant A, Auto Range)
-
Chromaticity 1
⊿
x
、⊿
y
:
±0.0040
2 cd/
㎡
or below
±0.0025
2 cd/
㎡
or above
Accuracy is guaranteed for measurable range only
-
Chromaticity 2
⊿
x
、⊿
y
:±0.01
For a combination of the standard light and following color glass
O-55
、
Y-48
、
A-73B
、
IRA-05
、
T-44
⊿
x
、⊿
y
:±0.03
For a combination of the standard light and following color glass
R-61
、
B-46
、
V-44
、
G-54