47
・
When command can not received
“ST”+CR+LF
“NO”+CR+LF
・
When measurement error
“ST”+CR+LF
“OK”+CR+LF
Start measurement
・
・
・
・
・
“NG”+CR+LF
4.3.8
STC Command
The instrument starts continuous measurement according to specified each setting after receive
this command. But the instrument does not send measured data after completing measurement.
This command is used for monitoring analogue output.
After first measurement, the instrument will send the status of continuous measurement.
Return value
0
:
Normal stop
1
:
Continuous measurement
2
:
Abnormal stop
PC
Command RD-80SA
“STC”+CR+LF
“OK”+CR+LF
Start measurement
・
・
・
・
・
“1”+CR+LF
“END”+CR+LF
※
The instrument does not send
“END” command until completing first measurement.