61
Repeatability (Standard Illuminance A, Auto range, SINGLE mode,
2σ)
Luminance
:
Within 2%
2 cd/
㎡
or below
With in 1%
2 cd/
㎡
or above
The specification is guaranteed for measurable range only
Chromaticity
:
xy
within 0.003
0.5 cd/m
2
- 2 cd/m
2
xy
within 0.002
2 cd/
㎡
or above
The specification is guaranteed for measurable range only
Temperature characteristics
±3% (0
℃
- 35
℃
based on the measured data at 20
℃
)
Humidity characteristics
±3% (60%RH or less, based on the measured data at 50%RH)
Measurement range
8 steps (Auto range or Manual range)
Adjustment volume
-
Over range, Under range adjustment function
Adjusts the over / under range level
Over range
:
About 2 - 3.4V (Default
:
3.3V)
Under range
:
About 0 - 1.2V (Default
:
0.03V)
-
Sensitivity of photo detector adjustment function
Adjusts the sensitivity of photo detector
Valid range
:
About 0 - 1.0V
Recommend
:
About 0.3 - 0.5V
(When raise the sensitivity, noise increase. Use recommended range)
Default applied voltage is as follow;
1.00V
±
0.005V at 200 cd/m
2
for standard illuminant A
-
Analogue output offset adjustment function
Adjusts analogue output offset
Valid range
:
About ±1V
Default
:
0.05V
(
in measurement of Dark
)