Adjustment Volume
-
Offset adjustment VR1
The Volume for output analogue voltage adjustment
Adjustable range : about ±1.0V
Default value
:
0.05V
-
Applied voltage to photo detector adjustment VR3
Adjusts the applied voltage to the photo detector (Photo Multiplier Tube)
Sensitivity of detector can be changed.
Adjustable range : about 0V - 1.0V
Stronger voltage makes sensitivity higher.
※
When you keep on setting high sensitivity, the life of detector shorten. Please set the voltage
from 0.3V to 0.5V.
-
Over range Voltage adjustment VR4
The voltage level which generate over range error can be adjusted.
Adjustable range : 2.0V - 3.4V
Default value
:
3.3V
-
Under range Voltage adjustment VR5
The voltage level which generate under range error can be adjusted.
Adjustable range :0V - 1.2V
Default value
:
0.03V
GND Terminal
Applied voltage
Check terminal
Over range
Voltage check terminal
Under range
Voltage adjustment VR
Applied voltage to photo detector
adjustment VR
Offset adjustment VR
Under range
Voltage adjustment VR
Under range
Voltage check terminal
TP1
TP3
TP4
TP5
VR1
VR3
VR4
VR5