29
3.2.2.2
Manual range setting
There are two type of the Manual range. This described how to set each Manual range setting.
1
Current measurement range setting appears
2
Set to [MANUAL] by pressing [ROTATION] and[CHANGE] switch
3
After pressing [ENTER] switch, following screen appears.
4
Press [ROTATION] switch to select one option from followings.
ALL RANGE
: The same measurement range is set to each OPEN / X2 / Y / Z filter.
Press [ROTATION] switch to increment measurement range until
appropriate measurement range.
EACH SELECT
: The measurement range are set to each OPEN / X2 / Y / Z filter
separately.
■
The procedure to set measurement range for each OPEN / X2 / Y / Z separately
1
When [EACH SELECT] appear , Press [CHANGE] switch.
2
After following screen appears, Set appropriate measurement range at each
filter by using
[CHANGE] and [ROTATION] switch. And press [SHIFT] switch to
move the cursor to another filter.
* AUTO / MANUAL *
AUTO
* MANUAL RANGE *
ALL RANGE 1
* MANUAL RANGE *
EACH SELECT
* MANUAL EACH *
OPEN R=1 X2 R=2
Y R=3 Z R=4
* AUTO / MANUAL *
MANUAL