57
4.5
Error Code List
Error cord obtained via [ERR] command and their details are as below.
Code
Details
Countermeasure
Message on display
E0000
Normal
E0001
E0015
E0016
E0017
E0018
System Error
Please contact your dealer or Topcon Technohouse
Advice us of the error code, the circumstance when the problem
happen in details.
If momentary interruption of power, let us know it also.
system
error
range system error
filter system error
sensor system error
motor system error
E0002
E0003
Error in internal memory
Internal memory error happen
Restart the instrument. If the instrument does not recover from the
error, Please contact your dealer or Topcon Technohouse. Advice us
of the error code, the circumstance when the problem happen in details.
If momentary interruption of power, let us know it also.
memory init error
memory access error
E0004
E0005
Failure of Photo detector calibration
Please contact your dealer or Topcon Technohouse.
Advise us of the error code, the circumstance when the problem
happen, and following points;
-
Error happen again after press [CALIB] switch ?
-
Error happen during measurement ?
-
Measurement data is correct ?
calibration error
calculation error
E0006
Receive parameter is abnormal
Check following points
- Parameter in sending is specified within valid value range?
- The number of parameter is correct?
parameter error