EF200F-C
Electrical connection
35
7.1.3
Connecting cable for remote version
Connecting cable
Standard cable
2 × 2 × 0.5 mm
2
(22 AWG) PVC cable with common shield (2 pairs, pair-
stranded)
1)
Flame resistance
According to DIN EN 60332-1-2
Oil-resistance
According to DIN EN 60811-2-1
Shielding
Galvanized copper-braid, opt. density approx.85 %
Cable length
30 m (98 ft)
Operating temperature
When mounted in a fixed position: ‒50 to +105
˚
C (‒58 to +221
˚
F);
when cable can move freely: ‒25 to +105
˚
C (‒13 to +221
˚
F)
1)
UV radiation may cause damage to the outer jacket of the cable. Protect the cable from exposure to sun as
much as possible.
Connecting cable (option "mass pressure-/temperature-compensated")
Order code for "Sensor version; DSC sensor; measuring tube"
Standard cable
[(3 × 2) + 1] × 0.34 mm
2
(22 AWG)PVC cable with common shield (3 pairs, pair-
stranded)
1)
Flame resistance
According to DIN EN 60332-1-2
Oil-resistance
According to DIN EN 60811-2-1
Shielding
Galvanized copper-braid, opt. density approx. 85%
Cable length
30 m (98 ft)
Operating temperature
When mounted in a fixed position: ‒50 to +105
˚
C (‒58 to +221
˚
F);
when cable can move freely: ‒25 to +105
˚
C (‒13 to +221
˚
F)
1)
UV radiation may cause damage to the outer jacket of the cable. Protect the cable from exposure to