EF200F-C
Electrical connection
43
4. Wire the connecting cable:
Terminal 1 = brown cable
Terminal 2 = white cable
Terminal 3 = green cable
Terminal 4 = red cable
Terminal 5 = black cable
Terminal 6 = yellow cable
Terminal 7 = blue cable
5. Connect the cable shield via the cable strain relief.
6. Tighten the screws for the cable strain relief using a torque in the range of
1.2 to 1.7 Nm.
7. Reverse the removal procedure to reassemble the connection housing.
Connecting transmitter via terminals
+
E
–
3 mm
1.
2.
3.
A0034173
1. Loosen the securing clamp of the electronics compartment cover.
2. Unscrew the electronics compartment cover.
3. Pull out the display module with a gentle rotational movement. To make it easier to
access the lock switch, attach the display module to the edge of the electronics
compartment.
8 mm
TX 10
4.
5.
A0034174
4. Loosen the locking screw of the transmitter housing.
5. Loosen the securing clamp of the transmitter housing.