10. In the Dynamic viscosity parameter, enter the viscosity of the fluid.
10.9.3 Gas applications
For accurate mass or corrected volume measurement, it is recommended to use the
pressure-/temperature-compensated sensor version.
If neither of these two options is possible, the pressure can also be entered as
a fixed value in the Fixed process pressure parameter.
Air
Select medium
Navigation:
Setup → Medium selection
1. Call up the Medium selection wizard.
2. In the Select medium parameter (→ See 10.4.3), select the Gas option.
3. In the Select gas type parameter (→ See 10.4.3), select the Air option.
The density is determined according to NEL 40.
4. Enter the value in the Relative humidity parameter (→ See 10.4.3).
The relative humidity is entered as a %. The relative humidity is converted
internally to absolute humidity and is then factored into the density calculation
according to NEL 40.
5. In the Fixed process pressure parameter (→ See 10.4.3), enter the value of the process
pressure present.
Configuring fluid properties
Navigation:
Setup → Advanced setup → Medium properties
6. Call up the Medium properties submenu.
7. In the Reference pressure parameter (→ See 10.5.1) enter the reference pressure for
calculating the reference density.
Pressure that is used as a static reference for combustion. This makes it possible
to compare combustion processes at different pressures.
8. In the Reference temperature parameter (→ See 10.5.1) enter the temperate for
calculating the reference density.