EF200F-C
Installation
29
Mounting the pressure measuring unit
4.
4 x
3.
5.
6 mm
5
6
A0035442
5
Siphon
6
Pressure measuring cell
3.
Damage to seal!
The seal is made of expanded graphite. It can therefore be used only once. If a
coupling is released, a new seal must be installed.
Use the additional seals provided. If necessary, these can be ordered as separate
spare parts at a later stage.
Insert the enclosed seal into the groove of the flange connection on the sensor side.
4. Align the flange connection on the pressure measuring unit and tighten the screws
by hand.
5. Tighten the screws with a torque wrench in three steps.
1. 10 Nm in criss-cross sequence
2. 15 Nm in criss-cross sequence
3. 15 Nm in circular sequence
►