EF200F-C
Diagnostics and troubleshooting
129
Diagnostic
number
Short text
Remedy instructions
Status
signal
[from the
factory]
Diagnostic
behavior
[from the
factory]
062
Sensor connection
defective
1. Check plug connections
2. Change pre-amplifier
3. Change DSC sensor
F
Alarm
082
Data storage
1. Check module connections
2. Contact service
F
Alarm
083
Memory content
1. Restart device
2. Restore S-Dat data
3. Change sensor
F
Alarm
114
Sensor leaky
Change DSC sensor
F
Alarm
122
Temperature sensor
defective
1. Check plug connections
2. Change pre-amplifier
3. Change DSC sensor
M
Warning
1)
170
Pressure cell connection
defective
1. Check plug connections
2. Replace pressure cell
F
Alarm
171
Ambient temperature
too low
Increase ambient temperature
S
Warning
172
Ambient temperature
too high
Reduce ambient temperature
S
Warning
173
Sensor range exceeded 1. Check process cond.
2. Increase system pressure
S
Warning
174
Pressure cell electronics
defective
Replace pressure cell
F
Alarm
175
Pressure cell deactivated Enable pressure cell
M
Warning
Diagnostic of electronic
242
Software incompatible
1. Check software
2. Flash or change main electronics
module
F
Alarm
252
Modules incompatible
1. Check if correct electronic modul
is plugged
2. Replace electronic module
F
Alarm
261
Electronic modules
1. Restart device
2. Check electronic modules
3. Change I/O Modul or main
electronics
F
Alarm
262
Module connection
1. Check module connections
2. Change electronic modules
F
Alarm
270
Main electronic failure
Change main electronic module
F
Alarm
271
Main electronic failure
1. Restart device
2. Change main electronic module
F
Alarm
272
Main electronic failure
1. Restart device
2. Contact service
F
Alarm
272
ECC settings faulty
F
Alarm
273
Main electronic failure
1. Emergency operation via display
2. Change main electronics
F
Alarm
275
I/O module defective
Change I/O module
F
Alarm
276
I/O module faulty
1. Restart device
2. Change I/O module
F
Alarm
276
I/O module faulty
F
Alarm
277
Electronics defective
1. Change pre-amplifier
2. Change main electronic module
F
Alarm
282
Data storage
1. Restart device
2. Contact service
F
Alarm