EF200F-C
Commissioning
101
10.5.6 Configuration management
After commissioning, you can save the current device configuration, copy it to
another measuring point or restore the previous device configuration.
You can do so using the Configuration management parameter and the related
options found in the Configuration backup display submenu.
Navigation
"Setup" menu → Advanced setup → Configuration backup display
Configuration backup display
Operating time
Last backup
Configuration management
Comparison result
Parameter overview with brief description
Parameter
Prerequisite
Description
User interface /
Selection
Factory setting
Operating time
‒
Indicates how long the device
has been in operation.
Days (d), hours (h),
minutes (m) and
seconds (s)
‒
Last backup
A local display is provided.
Indicates when the last data
backup was saved to the
display module.
Days (d), hours (h),
minutes (m) and
seconds (s)
‒
Configuration management
A local display is provided.
Select action for managing the
device data in the display
module.
• Cancel
• Execute backup
• Restore
• Duplicate
• Compare
• Clear backup data
Cancel
Comparison result
A local display is provided.
Comparison between present
device data and display
backup.
• Settings identical
• Settings not
identical
• No backup
available
• Backup settings
corrupt
• Check not done
• Dataset
incompatible
Check not done
Function scope of the "Configuration management" parameter
Options
Description
Cancel
No action is executed and the user exits the parameter.
Execute backup
A backup copy of the current device configuration is saved from the HistoROM
backup to the display module of the device. The backup copy includes the
transmitter data of the device.
►