EF200F-C
CommissionDiagnostics and troubleshootinging
127
12.3 Adapting the diagnostic information
12.3.1 Adapting the diagnostic behavior
Each item of diagnostic information is assigned a specific diagnostic behavior at the
factory. The user can change this assignment for specific diagnostic information in the
Diagnostic behavior submenu.
Expert → System → Diagnostic handling → Diagnostic behavior
A0014048-EN
Taking the example of the local display
You can assign the following options to the diagnostic number as the diagnostic behavior:
Options
Description
Alarm
The device stops measurement. The signal outputs and totalizers assume the defined
alarm condition. A diagnostic message is generated.
For local display with touch control: the background lighting changes to red.
Warning
The device continues to measure. The signal outputs and totalizers are not affected.
A diagnostic message is generated.
Logbook entry only The device continues to measure. The diagnostic message is displayed only in the
Event logbook submenu (Event list submenu) and is not displayed in alternation
with the operational display.
Off
The diagnostic event is ignored, and no diagnostic message is generated or entered.
0723-1
Diagnostic no. 044
/ ../Diagn. behavior
Diagnostic no. 801
Diagnostic no. 274
Warning