Operation
Settings
3-36
49iQPS Instruction Manual
Thermo Scientific
The Pressure Check screen is used to manually control the flow of reference
or sample gas through Cell B. This enables the pressure reading of Cell B,
with either sample or reference gas, to be determined. Pump pressure is
used to test the pump. Selecting any either of these will disturb the analog
outputs.
All pressure check screens function the same way. Therefore, the following
example of the Sample Pressure screen applies to all screens as well.
Home Screen>Settings>Health Check>Status and Alarms>Pressure
Check
Home Screen>Settings>Health Check>Status and Alarms>Pressure
Check>Sample Pressure
The Pressure Check screen contains the following information:
●
Sample Pressure:
Read only. Displays the pressure of the sample gas in
Cell B.
●
Reference Pressure:
Read only. Displays the pressure of the reference gas
in Cell B.
●
Pump Pressure:
Read only. Is used to test the pump.
Pressure Check
Содержание Ozone Primary Standard 49iQPS
Страница 1: ...49iQPS Instruction Manual Ozone Primary Standard 117434 00 1Apr2020...
Страница 2: ......
Страница 8: ......
Страница 31: ...Operation Instrument Display Thermo Scientific 49iQPS Instruction Manual 3 7...
Страница 142: ......
Страница 146: ...Maintenance Pump Rebuilding 4 4 49iQPS Instruction Manual Thermo Scientific Figure 4 2 Pump Rebuilding...
Страница 164: ...Servicing Fan Replacement 6 8 49iQPS Instruction Manual Thermo Scientific Figure 6 8 Replacing the Fan...
Страница 204: ...Servicing DMC Ozonator 6 48 49iQPS Instruction Manual Thermo Scientific Figure 6 55 Ozone DMC install pt 2...
Страница 208: ......
Страница 216: ......
Страница 234: ......
Страница 242: ......
Страница 254: ......
Страница 255: ......