Troubleshooting
Troubleshooting Guide
Thermo Scientific
49iQPS Instruction Manual
5-3
Problem
Possible Cause
Action
Flow, pressure or
vacuum are too low,
or: pump produces no
flow
Diaphragm or valve plates
are worn or dirt in the head
Rebuild or replace pump.
Connections or lines are
blocked
Let the pump run for a few minutes
pumping air.
An external valve is closed
or a filter blocked
Remove the external filter or valve to
see if the flow comes back into spec.
There is a leak at a
connector, in a line or in the
pump head
Find and repair leak.
Liquid (condensate) has
collected in the pump head
Clean and dry. Pump may need
replacement.
Pump cable not seated
Verify the pump cable is properly
seated into the Step POL board.
The cross-section of
pneumatic lines, or
connected components are
too small or they are
restricted
The pump is not designed for this
condition.
Alarm – board
communication
failure
Internal cables not
connected properly
Verify cables to DMC PCB are seated
properly. Replace DMC PCB if
necessary.
Alarm – power
supply
Internal cables not
connected properly
Verify cables to DMC PCB are seated
properly. Replace DMC PCB if
necessary.
Cell A or B frequency
(Photometer)
Light adjustment
Readjust Lamp setting from the
“Photometer Bench Settings” menu.
Defective detector
Interchange detectors at Photometer
DMC PCB connectors to determine if
detector is defective.
Cell A and B
frequency (Ozone)
Lamp supply
Cell A or B frequency
low or zero
Light adjustment
Readjust lamp setting from the
“Photometer Bench Settings” menu.
One cell excessively
contaminated
Clean cell.
Defective detector
Interchange detectors at photometer
DMC PCB connectors to determine if
detector is defective.
Cell A and B
frequency low or zero
Dirty cells
Clean cells.
Light
adjustment
Содержание Ozone Primary Standard 49iQPS
Страница 1: ...49iQPS Instruction Manual Ozone Primary Standard 117434 00 1Apr2020...
Страница 2: ......
Страница 8: ......
Страница 31: ...Operation Instrument Display Thermo Scientific 49iQPS Instruction Manual 3 7...
Страница 142: ......
Страница 146: ...Maintenance Pump Rebuilding 4 4 49iQPS Instruction Manual Thermo Scientific Figure 4 2 Pump Rebuilding...
Страница 164: ...Servicing Fan Replacement 6 8 49iQPS Instruction Manual Thermo Scientific Figure 6 8 Replacing the Fan...
Страница 204: ...Servicing DMC Ozonator 6 48 49iQPS Instruction Manual Thermo Scientific Figure 6 55 Ozone DMC install pt 2...
Страница 208: ......
Страница 216: ......
Страница 234: ......
Страница 242: ......
Страница 254: ......
Страница 255: ......