Servicing
DMC Ozonator
Thermo Scientific
49iQPS Instruction Manual
6-49
Use the following procedure to remove and replace the pressure regulator.
Equipment required:
Wrench
1.
Turn instrument OFF, unplug power cord, and remove the cover
(
Figure 2–1
).
2.
Disconnect 2x plumbing.
3.
Using a wrench, unfasten regulator nut. Remove as shown.
4.
Pull pressure regulator outwards.
Figure 6–56.
Pressure Regulator
Note
Make sure regulator notch is in the bracket slot before fastening
regulator nut. This will prevent the pressure regulator from rotating.
▲
5.
Replace pressure regulator and reassemble in reverse order.
Pressure Regulator