Servicing
49iQPS Photometer DMC
6-32
49iQPS Instruction Manual
Thermo Scientific
Use the following procedure to remove the Photometer DMC and replace
the board.
Equipment required:
Phillips drive, #2
1.
Turn instrument OFF, unplug power cord, and remove the cover
(
Figure 2–1
).
2.
Unplug DMC cable, top connector (photometer DMC board to
pressure/flow board).
3.
Unplug DMC cable, bottom connector (photometer DMC board to
ozonator DMC board).
4.
Disconnect 6X plumbing.
5.
Using a #2 Phillips drive, unfasten 6X captive hardware.
Figure 6–38.
Photometer DMC Removal
Photometer DMC
Removal