Troubleshooting
Troubleshooting Guide
Thermo Scientific
49iQPS Instruction Manual
5-5
Problem
Possible Cause
Action
If using a sample particulate filter,
make sure it is not blocked.
Disconnect sample particulate filter
from the sample bulkhead; if flow
increases, replace filter.
Alarm – Cell A
Intensity
Alarm – Cell B
Intensity
Pre-amp gain not set
properly
Check gain adjustment.
Defective photometer DMC
PCB board
Replace photometer DMC PCB board.
Alarm – Zero Check
(Option)
Instrument out of
calibration
Recalibrate instrument.
Check gas supply.
Perform a manual calibration.
Alarm – O
3
Concentration
Concentration has
exceeded range limit
Check to ensure range corresponds
with expected value. If not, select
proper range.
Concentration low
Check user-defined low set point, set
to zero.
Alarm – Lamp Heater
Current (Photometer)
Defective heater
Replace heater as needed.
Alarm – Lamp Temp
Sensor Short
(Photometer)
Defective sensor
Replace sensor as needed.
Alarm – Lamp Temp
Sensor Open
(Photometer)
Defective sensor
Replace sensor as needed.
Alarm – Bench Temp
Sensor Open
(Photometer)
Defective sensor
Replace sensor as needed.
Alarm – Bench Temp
Sensor Short
(Photometer)
Defective sensor
Replace sensor as needed.
Alarm – Lamp
Connection
(Photometer)
Defective lamp
Replace lamp as needed.
Alarm – Lamp Short
(Photometer)
Defective lamp
Replace lamp as needed.
Alarm – Lamp Heater
Current (Ozone)
Defective heater
Replace heater as needed.
Содержание Ozone Primary Standard 49iQPS
Страница 1: ...49iQPS Instruction Manual Ozone Primary Standard 117434 00 1Apr2020...
Страница 2: ......
Страница 8: ......
Страница 31: ...Operation Instrument Display Thermo Scientific 49iQPS Instruction Manual 3 7...
Страница 142: ......
Страница 146: ...Maintenance Pump Rebuilding 4 4 49iQPS Instruction Manual Thermo Scientific Figure 4 2 Pump Rebuilding...
Страница 164: ...Servicing Fan Replacement 6 8 49iQPS Instruction Manual Thermo Scientific Figure 6 8 Replacing the Fan...
Страница 204: ...Servicing DMC Ozonator 6 48 49iQPS Instruction Manual Thermo Scientific Figure 6 55 Ozone DMC install pt 2...
Страница 208: ......
Страница 216: ......
Страница 234: ......
Страница 242: ......
Страница 254: ......
Страница 255: ......