Servicing
49iQPS Photometer DMC
Thermo Scientific
49iQPS Instruction Manual
6-33
Use the following procedure to replace the board.
Equipment required:
Phillips drive, #2
1.
Unplug electrical, detector 1, detector 2, thermistor, and heater.
2.
Using a #2 Phillips drive, unfasten five 6-32 pan head screws.
Figure 6–39.
Photometer Board Replacement
3.
Replace board and reassemble in reverse order.
Photometer DMC
Board Replacement