Troubleshooting
Troubleshooting Guide
5-6
49iQPS Instruction Manual
Thermo Scientific
Problem
Possible Cause
Action
Alarm – Bench
Temperature (Ozone)
Dirty fan filter
Clean fan filter as needed.
Alarm – Lamp Temp
Sensor Short (Ozone)
Defective sensor
Replace sensor as needed.
Alarm – Lamp Temp
Sensor Open (Ozone)
Defective sensor
Replace sensor as needed.
Alarm – Lamp
Connection (Ozone)
Defective lamp
Replace lamp as needed.
Defective DMC board
Replace DMC board as needed.
Alarm – Lamp Short
(Ozone)
Defective lamp
Replace lamp as needed.
Reseat if needed.
Defective DMC PCB
Replace DMC board.
Содержание Ozone Primary Standard 49iQPS
Страница 1: ...49iQPS Instruction Manual Ozone Primary Standard 117434 00 1Apr2020...
Страница 2: ......
Страница 8: ......
Страница 31: ...Operation Instrument Display Thermo Scientific 49iQPS Instruction Manual 3 7...
Страница 142: ......
Страница 146: ...Maintenance Pump Rebuilding 4 4 49iQPS Instruction Manual Thermo Scientific Figure 4 2 Pump Rebuilding...
Страница 164: ...Servicing Fan Replacement 6 8 49iQPS Instruction Manual Thermo Scientific Figure 6 8 Replacing the Fan...
Страница 204: ...Servicing DMC Ozonator 6 48 49iQPS Instruction Manual Thermo Scientific Figure 6 55 Ozone DMC install pt 2...
Страница 208: ......
Страница 216: ......
Страница 234: ......
Страница 242: ......
Страница 254: ......
Страница 255: ......