Servicing
I/O Replacement
Thermo Scientific
49iQPS Instruction Manual
6-15
4.
During install, make sure to align cutout circular to keyway.
5.
Insert board downwards.
Figure 6–17.
I/O Replacement, Install
6.
Close arm. Make sure expansion bracket aligns to the inside of the
rectangular cutouts.
Figure 6–18.
I/O Replacement, Close Arm Alignment