![ThermoFisher Scientific Ozone Primary Standard 49iQPS Скачать руководство пользователя страница 54](http://html1.mh-extra.com/html/thermofisher-scientific/ozone-primary-standard-49iqps/ozone-primary-standard-49iqps_instruction-manual_1105615054.webp)
Operation
Settings
3-30
49iQPS Instruction Manual
Thermo Scientific
●
Cell A Concentration:
Displays the current concentration in cell A.
Alarm limits are not changeable.
●
Cell B Concentration:
Displays the current concentration in cell B.
Alarm limits are not changeable.
●
Cell A Intensity:
Displays the current intensity in cell A.
●
Cell B Intensity:
Displays the current intensity in cell B.
●
Lamp Temperature:
Displays the current lamp temperature. User
can adjust low and high alarm limits.
●
Lamp Temp Sensor Short:
Displays OK/Fail for lamp temperature
sensor short.
●
Lamp Temp Sensor Open:
Displays OK/Fail for lamp temperature
sensor open.
●
Lamp Current:
Displays the current lamp Current. User can adjust
low and high alarm limits.
●
Lamp Connection:
Displays OK/Fail for lamp connection.
●
Lamp Short:
Displays OK/Fail for lamp short.
●
Lamp Heater Current:
Displays the current lamp heater current.
●
Bench Temp Sensor Short:
Displays OK/Fail for bench temperature
sensor short.
●
Bench Temp Sensor Open:
Displays OK/Fail for bench temperature
sensor open.
●
Board Communication Failure:
Displays OK/Fail for board
communication status.
●
Power Supply:
Displays OK/Fail of power supplies. Power supply
goes red if any voltages are outside their limits. No voltage rows
ever get highlighted.
●
2.5 V Diagnostics:
Displays current voltage readings. Alarm
limits are not changeable.
●
2.5 V Ref. Diagnostics:
Displays current voltage readings. Alarm
limits are not changeable.
●
3.3 Volts:
Displays current voltage readings. Alarm limits are
not changeable.
●
5.0 V Diagnostics:
Displays current voltage readings. Alarm
limits are not changeable.
●
Lamp Power:
Displays current lamp power readings. Alarm
limits are not changeable.
Содержание Ozone Primary Standard 49iQPS
Страница 1: ...49iQPS Instruction Manual Ozone Primary Standard 117434 00 1Apr2020...
Страница 2: ......
Страница 8: ......
Страница 31: ...Operation Instrument Display Thermo Scientific 49iQPS Instruction Manual 3 7...
Страница 142: ......
Страница 146: ...Maintenance Pump Rebuilding 4 4 49iQPS Instruction Manual Thermo Scientific Figure 4 2 Pump Rebuilding...
Страница 164: ...Servicing Fan Replacement 6 8 49iQPS Instruction Manual Thermo Scientific Figure 6 8 Replacing the Fan...
Страница 204: ...Servicing DMC Ozonator 6 48 49iQPS Instruction Manual Thermo Scientific Figure 6 55 Ozone DMC install pt 2...
Страница 208: ......
Страница 216: ......
Страница 234: ......
Страница 242: ......
Страница 254: ......