![ThermoFisher Scientific Ozone Primary Standard 49iQPS Скачать руководство пользователя страница 88](http://html1.mh-extra.com/html/thermofisher-scientific/ozone-primary-standard-49iqps/ozone-primary-standard-49iqps_instruction-manual_1105615088.webp)
Operation
Settings
3-64
49iQPS Instruction Manual
Thermo Scientific
●
Down:
●
Detector 1 Manual Entry:
Shows current values for detector 1. If
detector cal cell is pressed, user can manually set the detector 1 cal
value.
●
Detector 2 Manual Entry:
Shows current values for detector 2. If
detector cal cell is pressed, user can manually set the detector 2 cal
value.
●
Auto Cal:
When Start is pressed, the auto-calibration process is
initiated. Please allow up to 5 minutes for calibration to complete.
User can stop calibration by pressing the Stop button.
●
End Cal:
When Stop is pressed, the auto calibration is interrupted
and the value does not change.
●
Default Values:
When pressed, the default values are saved.
Содержание Ozone Primary Standard 49iQPS
Страница 1: ...49iQPS Instruction Manual Ozone Primary Standard 117434 00 1Apr2020...
Страница 2: ......
Страница 8: ......
Страница 31: ...Operation Instrument Display Thermo Scientific 49iQPS Instruction Manual 3 7...
Страница 142: ......
Страница 146: ...Maintenance Pump Rebuilding 4 4 49iQPS Instruction Manual Thermo Scientific Figure 4 2 Pump Rebuilding...
Страница 164: ...Servicing Fan Replacement 6 8 49iQPS Instruction Manual Thermo Scientific Figure 6 8 Replacing the Fan...
Страница 204: ...Servicing DMC Ozonator 6 48 49iQPS Instruction Manual Thermo Scientific Figure 6 55 Ozone DMC install pt 2...
Страница 208: ......
Страница 216: ......
Страница 234: ......
Страница 242: ......
Страница 254: ......
Страница 255: ......