Operation
Settings
3-92
49iQPS Instruction Manual
Thermo Scientific
Home Screen>Settings>Instrument Settings>Advanced Instrument
Settings>49iQPS Plumbing>Update Plumbing
The Update Plumbing screen contains the following information:
●
One Sample Capillary:
If the instrument has 3 total capillaries (1
Sample capillary, 1 Reference Flow capillary, 1 Ozonator capillary) and
a Flow/Pressure DMC that has 2 absolute pressure sensors on it, the
user would choose One Sample Capillary.
●
Two Sample Capillary:
If the instrument has 4 total capillaries (2
Sample capillaries, 1 Reference Flow capillary, 1 Ozonator capillary)
and a Flow/Pressure DMC that has 3 absolute pressure sensors on it,
the user would choose Two Sample Capillaries.
Note
After selecting the number of capillaries, the user will be required to
reboot the instrument for the changes to the user interface to take effect.
▲
Содержание Ozone Primary Standard 49iQPS
Страница 1: ...49iQPS Instruction Manual Ozone Primary Standard 117434 00 1Apr2020...
Страница 2: ......
Страница 8: ......
Страница 31: ...Operation Instrument Display Thermo Scientific 49iQPS Instruction Manual 3 7...
Страница 142: ......
Страница 146: ...Maintenance Pump Rebuilding 4 4 49iQPS Instruction Manual Thermo Scientific Figure 4 2 Pump Rebuilding...
Страница 164: ...Servicing Fan Replacement 6 8 49iQPS Instruction Manual Thermo Scientific Figure 6 8 Replacing the Fan...
Страница 204: ...Servicing DMC Ozonator 6 48 49iQPS Instruction Manual Thermo Scientific Figure 6 55 Ozone DMC install pt 2...
Страница 208: ......
Страница 216: ......
Страница 234: ......
Страница 242: ......
Страница 254: ......
Страница 255: ......