Servicing
Peripherals Support Board and System Controller Board Replacement
6-16
49iQPS Instruction Manual
Thermo Scientific
Use the following procedure to replace the peripherals support board or
system controller board.
1.
Turn instrument OFF, unplug power cord, and remove the cover
(
Figure 2–1
).
2.
Pull tab out (two per board).
3.
Pull board out.
Figure 6–19.
Replacing the Peripherals Support Board
4.
Replace board and reassemble in reverse order.