Servicing
Optional Manifold Replacement
6-50
49iQPS Instruction Manual
Thermo Scientific
Use the following procedure to replace the manifold.
Equipment required:
Phillips drive, #2
Wrench
1.
Turn instrument OFF, unplug power cord, and remove the cover
(
Figure 2–1
).
2.
Using a wrench, unfasten three nuts. Remove the nuts, front and back
ferrules as shown from back panel.
Figure 6–57.
Replacing the Manifold pt 1
3.
Disconnect tubing.
4.
Using a #2 Phillips drive, unfasten four #8-32 socket cap head screws.
Optional
Manifold
Replacement