3-61
The Gas Units screen defines how the O
3
concentration reading is
expressed. The μg/m
3
, mg/m
3
, and g/m
3
gas concentration modes are
calculated using a standard pressure of 760 mmHg and a standard
temperature of 0 °C. The temperature can be changed in the compensation
screen.
Use the
and
buttons to select.
Home Screen>Settings>Measurement Settings>Gas Units
The Gas Units screen contains the following information:
●
ppb:
parts per billion.
●
ppm:
parts per million.
●
%:
percent.
●
μg/m3:
micrograms per meter cubed.
●
mg/m3:
milligrams per meter cubed.
●
g/m3:
grams per meter cubed.
Gas Units
Содержание Ozone Primary Standard 49iQPS
Страница 1: ...49iQPS Instruction Manual Ozone Primary Standard 117434 00 1Apr2020...
Страница 2: ......
Страница 8: ......
Страница 31: ...Operation Instrument Display Thermo Scientific 49iQPS Instruction Manual 3 7...
Страница 142: ......
Страница 146: ...Maintenance Pump Rebuilding 4 4 49iQPS Instruction Manual Thermo Scientific Figure 4 2 Pump Rebuilding...
Страница 164: ...Servicing Fan Replacement 6 8 49iQPS Instruction Manual Thermo Scientific Figure 6 8 Replacing the Fan...
Страница 204: ...Servicing DMC Ozonator 6 48 49iQPS Instruction Manual Thermo Scientific Figure 6 55 Ozone DMC install pt 2...
Страница 208: ......
Страница 216: ......
Страница 234: ......
Страница 242: ......
Страница 254: ......
Страница 255: ......