Operation
Calibration
Thermo Scientific
49iQPS Instruction Manual
3-9
The Adjust Background screen is used to make small adjustments to the
zero level. This adjustment should only be made when the 49iQPS is being
precisely matched to a NIST traceable O
3
calibrator. Normally, the
instrument does not require any background compensation.
The user can manually adjust the zero background by entering a value in
the Adjust Background button. The second button called Adjusted
Concentration shows what the new O
3
concentration would be based on
the changed zero background. Press the Calibrate button to save the
adjusted zero background value.
Home Screen>Calibration>Adjust Background
The Adjust Background screen contains the following information:
●
Adjust Background:
Allows the user to manually adjust the zero
background.
●
Adjusted Concentration:
Read only. Displays what the O
3
concentration
will become when the calibrate button is pressed.
●
Calibrate:
When pressed, updates the background value, making the
concentration go to zero.
Adjust Background
Содержание Ozone Primary Standard 49iQPS
Страница 1: ...49iQPS Instruction Manual Ozone Primary Standard 117434 00 1Apr2020...
Страница 2: ......
Страница 8: ......
Страница 31: ...Operation Instrument Display Thermo Scientific 49iQPS Instruction Manual 3 7...
Страница 142: ......
Страница 146: ...Maintenance Pump Rebuilding 4 4 49iQPS Instruction Manual Thermo Scientific Figure 4 2 Pump Rebuilding...
Страница 164: ...Servicing Fan Replacement 6 8 49iQPS Instruction Manual Thermo Scientific Figure 6 8 Replacing the Fan...
Страница 204: ...Servicing DMC Ozonator 6 48 49iQPS Instruction Manual Thermo Scientific Figure 6 55 Ozone DMC install pt 2...
Страница 208: ......
Страница 216: ......
Страница 234: ......
Страница 242: ......
Страница 254: ......
Страница 255: ......