Operation
Calibration
3-12
49iQPS Instruction Manual
Thermo Scientific
The Custom O
3
Levels screen lists six custom levels: 1, 2, 3, 4, 5, and 6.
Custom levels deal with ozonator control and configuration. A percentage
of 100% results in the maximum amount of ozone production. A
percentage of 0% results in no ozone production. The range of percentages,
however, are not linear.
Use the
and
buttons to move up and down and the
and
buttons to move left and right.
Home Screen>Calibration>Custom O3 Levels
The Custom O
3
Levels screen contains the following information:
●
Across:
●
O
3
Levels:
This column lists items associated with the O
3
concentration.
●
Value:
Displays the current value for each item.
●
Units:
Displays units for each item.
●
Down:
●
Level 1–6:
Displays user-defined concentration levels.
Custom O
3
Levels
Содержание Ozone Primary Standard 49iQPS
Страница 1: ...49iQPS Instruction Manual Ozone Primary Standard 117434 00 1Apr2020...
Страница 2: ......
Страница 8: ......
Страница 31: ...Operation Instrument Display Thermo Scientific 49iQPS Instruction Manual 3 7...
Страница 142: ......
Страница 146: ...Maintenance Pump Rebuilding 4 4 49iQPS Instruction Manual Thermo Scientific Figure 4 2 Pump Rebuilding...
Страница 164: ...Servicing Fan Replacement 6 8 49iQPS Instruction Manual Thermo Scientific Figure 6 8 Replacing the Fan...
Страница 204: ...Servicing DMC Ozonator 6 48 49iQPS Instruction Manual Thermo Scientific Figure 6 55 Ozone DMC install pt 2...
Страница 208: ......
Страница 216: ......
Страница 234: ......
Страница 242: ......
Страница 254: ......
Страница 255: ......