7
•
Maintenance and Service
Page 184
Vanquish Pumps (VH-P10-A-02) Operating Manual
7.15
Replacing the Doors
When
Damage of door
TIP
The maintenance procedures do not require that you remove the
doors. If this should ever be required for a specific reason or procedure,
follow the related steps in this section.
Parts required
Replacement door
Follow these steps
NOTICE
To avoid damage to the door hinges, be careful when
performing the following sequence of steps and do not apply force.
1. To remove a door, push the door upward while opening.
Open the door to a position in which the two hinges on the housing
are aligned in the grooves on the door. You can remove the door
only when the hinges are in the grooves.
Figure 71: Unhinging a door
2. Slightly tilt the door to the outside, away from the housing, and
remove the door.
3. To install the door, align the door with the hinges on the housing. Be
careful not to clamp tubing or capillaries between the door and the
enclosure.
No.
Description
1
Hinge on the housing
2
Reception groove on
the door
Содержание Vanquish VH-P10-A-02
Страница 1: ...Vanquish Pumps VH P10 A 02 Operating Manual 4820 4411 EN Revision 1 0 October 2017 thermoscientific ...
Страница 4: ...Contacting Us Page 4 Vanquish Pumps VH P10 A 02 Operating Manual ...
Страница 10: ...Contents Page 10 Vanquish Pumps VH P10 A 02 Operating Manual ...
Страница 30: ...2 Safety Page 30 Vanquish Pumps VH P10 A 02 Operating Manual ...
Страница 46: ...4 Unpacking Page 46 Vanquish Pumps VH P10 A 02 Operating Manual ...
Страница 80: ...5 Installation Page 80 Vanquish Pumps VH P10 A 02 Operating Manual ...
Страница 232: ...11 Appendix Page 232 Vanquish Pumps VH P10 A 02 Operating Manual ...
Страница 233: ...12 Index Vanquish Pumps VH P10 A 02 Operating Manual Page 233 12 Index ...
Страница 238: ...Thermo Fisher Scientific Inc 168 Third Avenue Waltham Massachusetts 02451 USA www thermofisher com ...