5.6.7
Connecting the Pump and Autosampler
Parts required
Capillary to connect the pump to the autosampler
Follow these steps
1. On the connecting capillary, remove the knurls.
2. Guide the connecting capillary through the guide holes in the
housings of both, the pump and autosampler.
Figure 26: Guide holes in the autosampler and pump (left: view from
autosampler; right: view from pump)
No.
Description
1
Capillary connecting the pump and autosampler:
1a
Capillary end connected to the autosampler injection valve
1b
Capillary end to be connected to pump outlet
2
Guide holes in autosampler housing and pump housing:
2a
Guide hole in autosampler housing (view from above)
2b
Guide hole in pump housing (view from below)
3. Reattach the knurls to the connection capillary.
4. In the pump, connect the capillary to the pump outlet.
5. In the autosampler, connect the capillary to the injection valve.
Содержание Vanquish VH-P10-A-02
Страница 1: ...Vanquish Pumps VH P10 A 02 Operating Manual 4820 4411 EN Revision 1 0 October 2017 thermoscientific ...
Страница 4: ...Contacting Us Page 4 Vanquish Pumps VH P10 A 02 Operating Manual ...
Страница 10: ...Contents Page 10 Vanquish Pumps VH P10 A 02 Operating Manual ...
Страница 30: ...2 Safety Page 30 Vanquish Pumps VH P10 A 02 Operating Manual ...
Страница 46: ...4 Unpacking Page 46 Vanquish Pumps VH P10 A 02 Operating Manual ...
Страница 80: ...5 Installation Page 80 Vanquish Pumps VH P10 A 02 Operating Manual ...
Страница 232: ...11 Appendix Page 232 Vanquish Pumps VH P10 A 02 Operating Manual ...
Страница 233: ...12 Index Vanquish Pumps VH P10 A 02 Operating Manual Page 233 12 Index ...
Страница 238: ...Thermo Fisher Scientific Inc 168 Third Avenue Waltham Massachusetts 02451 USA www thermofisher com ...