7
•
Maintenance and Service
Vanquish Pumps (VH-P10-A-02) Operating Manual
Page 175
7.11.2.2
Solvent Lines From Pump Inlet to Degasser
When
Damage or blockage of the solvent line
Parts and additional items required
•
Solvent line between pump inlet (solvent selector) and degasser
•
Solvent suitable for purging
Preparations
To flush out harmful substances, purge the pump with an appropriate
solvent.
Follow these steps
1. Remove the solvent line on the pump inlet and on the degasser.
2. Install the new solvent line.
3. To remove any air bubbles from the solvent lines, purge the pump.
7.11.2.3
Solvent Line From Degasser To Pump Head
When
Damage or blockage of the solvent line
Parts and additional items required
•
Solvent line between degasser and pump head inlet
•
Solvent suitable for purging
Preparations
To flush out harmful substances, purge the pump with an appropriate
solvent.
Follow these steps
1. Remove the solvent line on the degasser and on the inlet assembly
of the pump head.
2. Install the new solvent line.
3. To remove any air bubbles from the solvent lines, purge the pump.
Содержание Vanquish VH-P10-A-02
Страница 1: ...Vanquish Pumps VH P10 A 02 Operating Manual 4820 4411 EN Revision 1 0 October 2017 thermoscientific ...
Страница 4: ...Contacting Us Page 4 Vanquish Pumps VH P10 A 02 Operating Manual ...
Страница 10: ...Contents Page 10 Vanquish Pumps VH P10 A 02 Operating Manual ...
Страница 30: ...2 Safety Page 30 Vanquish Pumps VH P10 A 02 Operating Manual ...
Страница 46: ...4 Unpacking Page 46 Vanquish Pumps VH P10 A 02 Operating Manual ...
Страница 80: ...5 Installation Page 80 Vanquish Pumps VH P10 A 02 Operating Manual ...
Страница 232: ...11 Appendix Page 232 Vanquish Pumps VH P10 A 02 Operating Manual ...
Страница 233: ...12 Index Vanquish Pumps VH P10 A 02 Operating Manual Page 233 12 Index ...
Страница 238: ...Thermo Fisher Scientific Inc 168 Third Avenue Waltham Massachusetts 02451 USA www thermofisher com ...