7
•
Maintenance and Service
Vanquish Pumps (VH-P10-A-02) Operating Manual
Page 183
NOTICE
A firmware downgrade or incomplete firmware update may
result in loss of functionality or malfunctioning of the pump.
•
Do not interrupt communication between the Chromeleon software
and the pump at any time during the procedure.
•
At the beginning of the update process, a message appears showing
the firmware version currently installed in the pump and the version
that will be transferred from the Chromeleon software. If the
firmware installed in the pump is a later version than the version in
Chromeleon, cancel the download.
The firmware update may take several minutes.
•
Monitor the Audit Trail of the Chromeleon Instrument Configuration
Manager (or Server Configuration program) to see whether the
firmware update was successful or failed.
•
If the firmware update failed, turn the device off and on again and
repeat the firmware update.
•
If the firmware update fails repeatedly, contact Thermo Fisher
Scientific Technical Support for assistance.
3. After a successful firmware update, requalification of the pump may
be required. See the release notes for a recommendation.
Содержание Vanquish VH-P10-A-02
Страница 1: ...Vanquish Pumps VH P10 A 02 Operating Manual 4820 4411 EN Revision 1 0 October 2017 thermoscientific ...
Страница 4: ...Contacting Us Page 4 Vanquish Pumps VH P10 A 02 Operating Manual ...
Страница 10: ...Contents Page 10 Vanquish Pumps VH P10 A 02 Operating Manual ...
Страница 30: ...2 Safety Page 30 Vanquish Pumps VH P10 A 02 Operating Manual ...
Страница 46: ...4 Unpacking Page 46 Vanquish Pumps VH P10 A 02 Operating Manual ...
Страница 80: ...5 Installation Page 80 Vanquish Pumps VH P10 A 02 Operating Manual ...
Страница 232: ...11 Appendix Page 232 Vanquish Pumps VH P10 A 02 Operating Manual ...
Страница 233: ...12 Index Vanquish Pumps VH P10 A 02 Operating Manual Page 233 12 Index ...
Страница 238: ...Thermo Fisher Scientific Inc 168 Third Avenue Waltham Massachusetts 02451 USA www thermofisher com ...